  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 闫威华张立国张钊肖志刚

【Author】 YAN Wei-Hua 1 ZHANG Li-Guo 2 ZHANG Zhao 1 XIAO Zhi-Gang 1;1) 1 Department of Physics,Tsinghua University,Beijing 100084,China 2 Institute of Nuclear Energy and Technology,Tsinghua University,Beijing 100084,China

【机构】 Department of Physics,Tsinghua UniversityInstitute of Nuclear Energy and Technology,Tsinghua University

【摘要】 The application of a high purity Germanium(HPGe) γ spectrometer in determining the fuel element burnup in a future reactor is studied.The HPGe detector is exposed by a 60Co source with varying irradiation rate from 10×103s-1 to 150×103s-1 to simulate the input counting rate in real reactor environment.A137 Cs and a152 Eu source are positioned at given distances to generate a certain event rate in the detector with the former being proposed as a labeling nuclide to measure the burnup of a fuel element.It is shown that both the energy resolution slightly increasing with the irradiation rate and the passthrough rate at high irradiation level match the requirement of the real application.The influence of the background is studied in the different parameter sets used in the specially developed procedure of background subtraction.It is demonstrated that with the typical input irradiation rate and 137Cs intensity relevant to a deep burnup situation,the precision of the 137Cs counting rate in the current experiment is consistently below 2.8%,indicating a promising feasibility of utilizing an HPGe detector in the burnup measurement in future bed-like reactors.

【Abstract】 The application of a high purity Germanium(HPGe) γ spectrometer in determining the fuel element burnup in a future reactor is studied.The HPGe detector is exposed by a 60Co source with varying irradiation rate from 10×103s-1 to 150×103s-1 to simulate the input counting rate in real reactor environment.A137 Cs and a152 Eu source are positioned at given distances to generate a certain event rate in the detector with the former being proposed as a labeling nuclide to measure the burnup of a fuel element.It is shown that both the energy resolution slightly increasing with the irradiation rate and the passthrough rate at high irradiation level match the requirement of the real application.The influence of the background is studied in the different parameter sets used in the specially developed procedure of background subtraction.It is demonstrated that with the typical input irradiation rate and 137Cs intensity relevant to a deep burnup situation,the precision of the 137Cs counting rate in the current experiment is consistently below 2.8%,indicating a promising feasibility of utilizing an HPGe detector in the burnup measurement in future bed-like reactors.

【基金】 Supported by National Science and Technology Major Project(ZX06901);National Natural Science Foundation of China (10975083);Tsinghua University Initiative Scientific Research Program
  • 【文献出处】 Chinese Physics C ,中国物理C , 编辑部邮箱 ,2012年11期
  • 【分类号】TL817.2
  • 【被引频次】5
  • 【下载频次】23
节点文献中: 

本文链接的文献网络图示:

本文的引文网络