FTIR光谱基线漂移快速校正的DSP实现

Relization of FTIR Baseline Drift Fast Correction Based on DSP

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 马恒张记龙赵冬娥王志斌陈媛媛

【Author】 MA Heng1,ZHANG Jilong1,2,ZHAO Dong’e1,2,WANG Zhibin1,2,CHEN Yuanyuan1(1.Engineering Technology Research Center of Shanxi Province for Opto-Electronic Information and Instrument,Taiyuan 030051,China;2.Key Laboratory of Instrumentation Science & Dynamic Measurement(North University of China),Ministry of Education,Taiyuan 030051,China)

【机构】 山西省光电信息与仪器工程技术研究中心中北大学仪器科学与动态测试教育部重点实验室

【摘要】 针对傅里叶变换红外光谱仪(FTIR)在长时间开放光程条件下进行测量时光谱基线的漂移,采用多项式迭代拟合逼近的方法,针对3种不同类型的基线漂移光谱进行校正,分别在MATLAB环境下及DSP中实现,校正后的光谱数据采用均方根误差R作为评价标准,分别达到了4.21×10-6和5.38×10-6,校正时间分别消耗0.935 0 s和0.778 3 s,结果表明该方法自动化程度高,校正效果好,速度快,可以为基于DSP实现系统小型化、集成化的光谱处理系统提供理论基础。

【Abstract】 Fourier transform infrared spectroscopy(FTIR) measurements in the long open-path conditions,FTIR spectra baseline will drift to influence the spectral information,changing the spectral shape and affect the accuracy of gas analysis.Therefore,the baseline correction of the measured spectral information is very important preprocessing step,providing a good data base for the next gas qualitative and quantitative analysis.For baseline drift using the approximation method of polynomial iterative fitting spectra of three different types of baseline drift correction,respectively under MATLAB and DSP enviroment.After correction for the spectral data to evaluate the root mean square error(R) respectively to 4.21×10-6 and 5.38×10-6,the correction time consumption are 0.935 0 s and 0.778 3 s.The results show that the method owns advantages of high degree of automation,the correction effect and high calculating speed,providing a theoretical basis for the miniaturization of integrated spectral processing system based on DSP.

【基金】 国家自然科学基金仪器专项基金项目(61127015);国家自然基金主任基金项目(61040062);国际科技合作项目(S2012ZR0247);山西省国际科技合作项目(2010081038)