节点文献

离子色谱法测定饮用水中高氯酸盐含量研究

Determination of perchlorate in drinking water by Ion chromatography

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 颜金良王立

【Author】 YAN Jin-ling,WANG Li(Ningbo Key Laboratory of Poison Research and Control,Ningbo Municipal Center for Disease Control and Prevention,Ningbo 315010,China)

【机构】 宁波市疾病预防控制中心宁波市毒物研究与控制重点实验室

【摘要】 目的:建立一种简便快速梯度淋洗离子色谱法测定饮用水中高氯酸盐含量。方法:选用以高容量、强亲水性IonPac AS23阴离子分析柱,0 mmol/L~40 mmol/LKOH淋洗液,流速为1.0 m l/m in,电导检测器,样品经0.45μm滤膜过滤后直接进样。结果:本法相关性好(r=0.9998),相对标准偏差(RSD)为2.15%~5.99%,样品加标平均回收率为87.9%~101%,最低检出量为3.0μg/L。结论:该法操作简单、快速、准确、灵敏、适用于大批水样分析。

【Abstract】 Objective:To develop an assay for determination of perchlorate in drinking water by ion chromatography(IC).Methods: The method was based on the ion chromatographic separation under the condition of isocratic eluent with 0 mmol/L~40 mmol/L KOH at a flow rate of 1.0 ml/min by using of the IonPAC AS23 analytical column and suppressed conductivity detector.Results:It was showed that this method had a good relativity(r=0.9998) and a good analytic precision(RSD 2.15%~5.99%).The average recoveries were 87.9%~101%,and the detection limit were 3.0 μg/L.Conclusion:Its advantages are simple operation,quickness,high sensitivity and precise results.This developed method was fast,sensitive,and accurate,it can be applied for the analysis of water samples.

【关键词】 高氯酸盐离子色谱
【Key words】 PerchlorateIon chromatographyDrinking water
【基金】 宁波市重点实验室建设基金(2008A22002)
  • 【文献出处】 中国卫生检验杂志 ,Chinese Journal of Health Laboratory Technology , 编辑部邮箱 ,2011年06期
  • 【分类号】O657.75;R123.1
  • 【被引频次】12
  • 【下载频次】310
节点文献中: 

本文链接的文献网络图示:

本文的引文网络