Comparison of Intraocular Pressures at Different Points in Human’s Cornea before and after Laser in situ Keratomileusis with Tono-Pen Tonometer
【摘要】 In order to explore the difference of intraocular pressure(IOP) at different points of cornea before and after laser in situ keratomileusis(LASIK),IOP was measured by Tono-Pen Tonometer at central cornea,pericentral cornea and limbus respectively and analyzed statistically.After LASIK,IOP was dropped significantly at central cornea and pericentral cornea(P<0.05),while no statistically significant change occurred at limbus(P>0.05).There was no statistically significant difference in IOP at different points before LASIK(F=0.110,P=0.896),but statistically significant difference was found after LASIK(F=7.375,P=0.001).It was suggested that reliable IOP after LASIK could be obtained from the limbus by Tono-Pen tonometer.
【Abstract】 In order to explore the difference of intraocular pressure(IOP) at different points of cornea before and after laser in situ keratomileusis(LASIK),IOP was measured by Tono-Pen Tonometer at central cornea,pericentral cornea and limbus respectively and analyzed statistically.After LASIK,IOP was dropped significantly at central cornea and pericentral cornea(P<0.05),while no statistically significant change occurred at limbus(P>0.05).There was no statistically significant difference in IOP at different points before LASIK(F=0.110,P=0.896),but statistically significant difference was found after LASIK(F=7.375,P=0.001).It was suggested that reliable IOP after LASIK could be obtained from the limbus by Tono-Pen tonometer.
【Key words】 Tono-Pen tonometer; laser in situ keratomileusis; cornea; intraocular pressure;
- 【文献出处】 Journal of Huazhong University of Science and Technology(Medical Sciences) ,华中科技大学学报(医学英德文版) , 编辑部邮箱 ,2011年01期
- 【分类号】R779.63
- 【被引频次】3
- 【下载频次】13