节点文献

MC-ICP-MS测量Ru同位素丰度比值的质量歧视校正

Correction of Mass Discrimination for Measuring Ru Isotope Abundance by MC-ICP-MS

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 常志远张继龙姜小燕赵永刚

【Author】 CHANG Zhi-yuan,ZHANG Ji-long,JIANG Xiao-yan,ZHAO Yong-gang(Department of Radiochemistry,China Institute of Atomic Energy,Beijing 102413,China)

【机构】 中国原子能科学研究院放射化学研究所

【摘要】 采用MC-ICP-MS测量Ru同位素丰度比时,存在较大的质量偏倚。利用IsoprobeMC-ICP-MS测量了RuCl3中Ru同位素丰度比值,并利用幂、指数以及通用幂校正规律(GPL)对Ru同位素丰度比值测量的质量偏倚进行了校正。结果表明,以100Ru/102Ru、104Ru/102Ru作为内标,采用GPL校正质量偏倚,得到的Ru同位素丰度更接近于天然值,与天然值的偏差在(20~650)×10-6之间。

【Abstract】 The large mass bias was corrected during the measurement of Ru isotope abundance by multicollector inductively coupled plasma mass spectrometer(MC-ICP-MS).A natural RuCl3 sample was determined by isoprobe MC-ICP-MS.The power law,exponential law and generalised power law(GPL)were adopted to correct the mass discrimination.The results show that if 100Ru/102Ru,104Ru/102Ru ratio are adopted to be the normalization,the mass bias can be well corrected by GPL.The correcting values of isotopic abundance ratios of Ru agree with the natural values at the level of(20—650)×10-6.

  • 【文献出处】 质谱学报 ,Journal of Chinese Mass Spectrometry Society , 编辑部邮箱 ,2010年02期
  • 【分类号】O657.63
  • 【被引频次】1
  • 【下载频次】176
节点文献中: 

本文链接的文献网络图示:

本文的引文网络