吉林省部分地区的X射线诊断设备质控检测结果与分析
Analysis of results of inspection for quality control of X-ray machine in parts of Jilin Province
【摘要】 目的:对部分医疗单位的X射线诊断设备性能检测和评价。方法:按照国家现行的设备质量检测规范要求,对三种类型的71台设备逐项检测。结果:全部项目合格率最高的是透视设备为51.2%、最低的是CR系统为8.4%;结论:总体看全部项目合格率为43.8%,透视设备检测合格率要好于其它。
【Abstract】 Objective: To survey the quality control of X-ray machine in some institutions.Methods: A total of 71 machines of 3 types were inspected according to the current national standards. Results: The qualified rate was the highest for perspective machine (51.2 %) and the lowest one for CR system (8.4 %).Conclusion: The overall qualified rate is 43.8 %.The qualified rate of perspective machine is higher than that of other types of machine.
【关键词】 X射线诊断设备;
质量控制检测;
合格率;
【Key words】 X-ray diagnostic instrument; Quality control examination; Qualified rate;