超声波体外充电的实验研究

Experimental Study on Ultrasound in Vitro Charge

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 谢育梅曾国洪冯琴昌温志浩陈玉芳梁召云黄德群陈雄英许少波

【Author】 XIE Yumei1,ZENG Guohong1,FENG Qinchang2,WEN Zhihao3,CHEN Yufang2,LIANG Zhaoyun2,HUANG Dequn2,CHEN Xiongying1,XU Shaobo2 ( 1Guangdong Cardiovascular Institute,Guangdong Academy of Medical Sciences,Guangdong General Hospital,Guangzhou 510100,China; 2Guangdong Medical Instruments Institute,Guangzhou 510500,China; 3Guangdong Food and Drug Vocational College,Guangzhou 510520,China)

【机构】 广东省心血管病研究所广东省医学科学院广东省人民医院广东省医疗器械研究所广东食品药品职业学院

【摘要】 目的探讨以超声波作为能量载体实现对心脏起搏器电池进行体外无创充电的可行性及有效性。方法采用超声波作为能量载体,通过压电陶瓷、耦合匹配、后级整理等电路进行能量的无线发送与接收,并加入磁耦合等补偿电路,提高超声波的能量传输效率。以水和成年猪的肌肉组织为不同的传输介质,超声波充电后产生的电压和电流作对比研究。结果超声波发生器在体外充电过程中的实际工作参数为:P=10.0W,f=(60~70)KHz。连接不同的阻抗负载电阻的三套电路,在负载小于100KΩ时,充电电路的输出电流随负载的降低而变化不大,输出电压随负载降低而降低;在负载大于1MΩ时,充电电路的输出电压随负载的增大而变化不大,输出电流随负载增大而降低;在负载于(100KΩ~1MΩ)范围内,充电电路输出功率处于最高的范围,在(0.5~1)mW范围内,随负载的增大而变化不大。结论通过超声波发生器在体外通过肌肉组织对起搏器电路进行无创式充电是可行的。

【Abstract】 Objective To explore the feasibility and effectiveness of the ultrasound,as an energy carrier,in achieving non-invasive battery charging of cardiac pacemakers. Methods The ultrasound was used as an energy carrier,sending and receiving energy wirelessly through circuit such as the piezoelectric ceramics,coupling matching,etc.,and meanwhile adding compensating circuit like magnetic coupling to enhance the transmission efficiency of ultrasonic energy. Pure distilled water and older pigs’ muscular tissue were practiced as two different transmission medium,the voltages and currents after charging the ultrasonic were comparatively studied. Results During the in vitro charging process of ultrasonic generator the actual operating limits were:P =10.0 W,f = (60 ~ 70) KHz. Three sets of circuits,connecting different impedance load resistance. When the load was less than 100 KΩ,the output current of the charging circuit changed little with the decrease of the load and the output voltage was reduced as the load decreases; when the load was more than 1 MΩ,the output voltage of the charging circuit changed little with the increase of the load and the output current was reduced as the load increases; when the load was between 100 KΩ to 1 MΩ,the output power of the charging circuit was at the highest range and it had little changes with the increase of the load when the load was between 0.5 mW to 1 mW. Conclusion It is feasible to charge the circuit of pacemakers in a non-invasive way through the muscle tissue by the ultrasonic generator in vitro.

【基金】 广东省科技攻关项目资助(基金编号:2006B12901009)
  • 【文献出处】 临床医学工程 ,Clinical Medicine & Engineering , 编辑部邮箱 ,2010年04期
  • 【分类号】R318.6
  • 【被引频次】7
  • 【下载频次】203