不同眼压计对可疑青光眼患者眼压测量的研究

Analysis of the Tonometer Type on IOP Measurement in Dubious Glaucoma

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 彭丽辉曾庆梅林先轩刘韶瑞骆荣江朱文珲

【Author】 PENG Li-hui,ZENG Qing-mei,LIN Xian-xuan,et al.Department of Ophthalmology,the First Affiliated Hospital of Sun Yat-Sen University,Guangzhou 510080,Guangdong,China

【机构】 中山大学附属第一医院眼科

【摘要】 目的比较Goldmann压平眼压计(GAT)、Schiotz眼压计和非接触眼压计(NCT)对青光眼排查患者测得的眼压值,探讨眼压计类型对患者眼压值的影响。方法对43例可疑青光眼排查患者(67只眼,34~58岁),进行24小时眼压曲线测量。先在卧位下使用压陷眼压计进行测量,然后在坐位下分别使用压平和非接触眼压计进行测量。采用方差分析比较不同测量方法所获得的眼压值。结果 Schiotz眼压计、GAT及NCT测得的眼压平均值分别为(24.58±9.76)、(22.13±8.25)、(18.51±7.48)mmHg(1mmHg=0.133kPa)。所测得的眼压值,GAT较NCT高1.7~3.1mmHg,Schiotz眼压计较NCT高3.0~6.4mmHg。三种测量方法测得的峰值眼压超过21mmHg的患者分别为36,27,19例,使用非接触眼压计可能漏诊青光眼的患者。结论使用GAT、Schiotz眼压计和NCT所得出的眼压曲线有所不同,使用NCT对靶眼压附近的眼压难以作出准确的判断。

【Abstract】 Objective To investigate the different tonometer’s influence on the measuring value of intraocularpressure(IOP) and to discuss the influences of different tonometer on the judgment of intraocularpressue(IOP) readings in patients with dubious glaucoma.Methods The diurnal IOP was measured every two hours in 24 hours using Schiotz tonometer in supine position,by goldmann applanation tonometer(GAT) and non-contact tonometer(NCT) in sitting position.IOP curve was drawn in 43 patients with dubious glaucoma.The IOP readings measured by different tonometer and posture in various time points were compared.Results The peak IOP value by Schiotz tonometer,GAT and by NCT was(24.58±9.76) mm Hg,(22.13±8.25) mm Hg and(18.51±7.48) mm Hg,respectively.The supine IOP readings by Schiotz tonometer and GAT were higher than those of NCT in various time points.The peak supine IOP value by Schiotz tonometer was over 21 mm Hg in 36 patients,and that by GAT was in 27 patients and 19 patients by NCT.Some patients with glaucoma were missed diagnosis by NCT.Conclusion The tonometer with dubious glaucoma in measuring IOP had certain influence on the IOP readings.The IOP value from NCT had the limiting accuracy.

【关键词】 眼压青光眼排查眼压计
【Key words】 Intraocular pressureGlaucoma test individuallyTonometer
  • 【文献出处】 中华全科医学 ,Chinese Journal of General Practice , 编辑部邮箱 ,2010年12期
  • 【分类号】R775.04
  • 【被引频次】7
  • 【下载频次】132
节点文献中: 

本文链接的文献网络图示:

本文的引文网络