Optimum condition of efficiency functions for HPGe γ-ray detectors in the 121-1408 keV energy range

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 陈志林宋国旸穆龙王和义邢世雄郭洪波廖振兴陈平华生

【Author】 CHEN Zhi-Lin SONG Guo-Yang MU Long WANG He-Yi XING Shi-Xiong GUO Hong-Bo LIAO Zhen-Xing CHEN Ping HUA Sheng Institute of Nuclear Physics and Chemistry,China Academy of Engineering Physics,Mianyang 621900,China

【机构】 Institute of Nuclear Physics and Chemistry,China Academy of Engineering Physics

【摘要】 The optimum condition of three commonly used functions in the Genie 2000 γ spectra analysis software have been studied in the 121-1408 keV energy range.The three functions are applied for fitting the full-energy peak effciency of the HPGe gammaray detector.A detailed procedure to obtain the optimum condition is described.The HPGe detector is calibrated at 11 cm by three radioactive sources of point form(152Eu,137Cs,60Co) providing 11 energy peaks.After data processing,results shows that the three functions used in the Genie 2000 gamma spectra analysis software fit best at orders 3-5.Lastly the standard radioactive source 133Ba is chosen to validate the results.Differences between the standard activity of 133Ba and the result obtained from the fitting functions are below 1.5%.Therefore the optimum orders of the three functions used in the Genie 2000 γ spectra analysis software are 3-5 with the 11 energy peaks.

【Abstract】 The optimum condition of three commonly used functions in the Genie 2000 γ spectra analysis software have been studied in the 121-1408 keV energy range.The three functions are applied for fitting the full-energy peak effciency of the HPGe gamma-ray detector.A detailed procedure to obtain the optimum condition is described.The HPGe detector is calibrated at 11 cm by three radioactive sources of point form(152Eu,137Cs,60Co) providing 11 energy peaks.After data processing,results shows that the three functions used in the Genie 2000 gamma spectra analysis software fit best at orders 3-5.Lastly the standard radioactive source 133Ba is chosen to validate the results.Differences between the standard activity of 133Ba and the result obtained from the fitting functions are below 1.5%.Therefore the optimum orders of the three functions used in the Genie 2000 γ spectra analysis software are 3-5 with the 11 energy peaks.

【关键词】 HPGefit functioneffciency calibrationGenie 2000
【Key words】 HPGefit functioneffciency calibrationGenie 2000