节点文献
Irradiation methods and nozzle design for the advanced proton therapy facility
【摘要】 A conception design of the Advanced Proton Therapy Facility(APTF) has been carried out.The system intentionally employs a slow-cycling synchrotron with a maximum energy of 250 MeV,two rotating gantries and two fixed beam nozzles for the treatment.In this paper,we try to compare the strength and weaknesses between the two treatment methods:the beam spreading and the pencil beam scanning.The application of the pencil beam scanning method and the double-scattering method together with the related nozzle design at APTF is also given.The simulation results of employing the double-scattering method have been given during the preliminary design.
【Abstract】 A conception design of the Advanced Proton Therapy Facility(APTF) has been carried out.The system intentionally employs a slow-cycling synchrotron with a maximum energy of 250 MeV,two rotating gantries and two fixed beam nozzles for the treatment.In this paper,we try to compare the strength and weaknesses between the two treatment methods:the beam spreading and the pencil beam scanning.The application of the pencil beam scanning method and the double-scattering method together with the related nozzle design at APTF is also given.The simulation results of employing the double-scattering method have been given during the preliminary design.
【Key words】 hadron therapy; irradiation method; nozzle; pencil scanning; spreading system; gantry;
- 【文献出处】 中国物理C ,Chinese Physics C , 编辑部邮箱 ,2010年04期
- 【分类号】TL543
- 【被引频次】1
- 【下载频次】45