优化牙釉质辐射剂量重建方法中的电子顺磁共振谱拟合程序

Optimization of Electron Paramagnetic Resonance Spectrum Fitting Procedures in Radiation Dose Reconstruction Using Enamel

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 周莉张文艺于奡苑淑渝戴光复张良安

【Author】 ZHOU Li1, ZHANG Wen-yi2, YU Ao3, YUAN Shu-Yu2, DAI Guang-fu2, Bai Sen1, ZHANG Liang-an2. 1. Division of Physics Center, West China Hospital, Sichuan University, Chengdu, Sichuan 610041, P.R. China; 2. Institute of Radiation Medicine, Chinese Academy of Medical Sciences & China Union Medical College, Tianjin 300192, P. R. China; 3. Central Laboratory, Nankai University, Tianjin 300071, P. R. China

【机构】 四川大学华西医院放射物理技术中心中国医学科学院中国协和医科大学放射医学研究所南开大学中心实验室

【摘要】 目的优化牙釉质电子顺磁共振(EPR)谱拟合程序。方法将牙釉质EPR谱拟合程序中的RIS-BGS偏移量设置为定参,取优化值;增加手动拟合功能,可以将参数初始值直接代入高斯模型得到拟合谱。结果将牙釉质EPR谱拟合程序中RIS-BGS偏移量设置为定参后,其优化值为0.529,减少了变参数量;采用手动拟合功能有助于调整参数初始值;在自动拟合程序对于个别样品谱不能给出正确结果情况时,仍然可以通过手动拟合来得到一个近似的拟合结果。结论优化了EPR谱拟合程序的应用过程。

【Abstract】 Objective To optimize the enamel electron paramagnetic resonance (EPR) spectrum fitting procedure. Methods RIS-BGS offset was set as a determined parameter in enamel EPR spectrum fitting program, and an optimized value was given for RIS-BGS offset. A manual fitting function was designed that can substite the initutial value into the Gaussian model function directly. Results RIS-BGS offset was set as a determined parameter, the optimized value of which was 0.529. The number of parameters decreased; the manual fitting function maked it easier to select initial values for parameters. When the auto fitting program failed to give a correct result for some EPR spectrum, a fitting result could still be obtained with the manual fitting method. Conclusion The application of the spectrum fitting procedure can be optimized.

【关键词】 牙釉质电子顺磁共振
【Key words】 Tooth enamelElectron paramagnetic resonance