节点文献
Z-Pinch软X射线探测器件的同步辐射标定
Synchrotron Radiation Calibration for Soft X-ray Detector(In Chinese)
【摘要】 介绍了利用同步辐射装置3W1B光站软X光源标定X光胶片、闪烁体、透射光栅的实验。实验中标定了闪烁体的能量响应、透射光栅的透过率和衍射效率,标定出软X光胶片黑密度曲线和能量响应曲线。实验结果可用于快过程Z-Pinch X光辐射特性的诊断,如软X射线功率仪、光栅谱仪、针孔相机、一维时空成像系统等测试项目,保证了Z-Pinch实验数据的精度。
【Abstract】 The calibration experiments were carried out to X-ray film,scintillator and transmission grating by employing the soft X-ray station at 3W1B beam-line in Beijing synchrotron Radiation Facility.The experiments presented the black intensity curve and energy response curve of soft X-ray film.And the experimental results can be used in diagnosis of X-ray radiation characterization of Z-pinch,such as in the measurement of soft X-ray Power Meter,grating spectrometer,pinhole camera and one-dimension imaging system which can ensure precision of Z-pinch results.
- 【文献出处】 中国核科技报告 ,China Nuclear Science and Technology Report , 编辑部邮箱 ,2009年01期
- 【分类号】O434.12
- 【下载频次】59