节点文献

TTC-脱氢酶活性测定中标准曲线的影响因素研究

Study on influencing factors for standard curve of TTC-Dehydrogenase activity measurement

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 赵连梅池勇志张春青

【Author】 ZHAO Lian-mei,CHI Yong-zhi,ZHANG Chun-qing(Department of Environmental and Municipal Engineering,Tianjin Institute of Urban Construction,Tianjin 300384,China)

【机构】 天津城市建设学院环境与市政工程系

【摘要】 此文采用722S型可见分光光度计,以连二亚硫酸钠为还原剂,对传统TTC-脱氢酶活性测定法中影响标准曲线制作的因素进行了研究,确定了最佳测定条件为:最大吸收峰在475nm处;连二亚硫酸钠的最佳投量为5.0mg/5mL;TTC溶液的保存及整个实验过程应避光。最佳反应时间为30min。

【Abstract】 In this paper the factors which influence the standard curve of determination are investigated by using 722S type visible pectrophotometer and sodium hydrosulfite as reductant in the measure method conventional TTC-Dehydrogenase activity.The optimum conditions are as follows:the maximum absorption peak is at the point of 475 nm;the best dosage of sodium dithionite is 5.0 mg/5mL ;preservation of TTC solution and the whole test process should be in darkness;the optimum reaction time is 30 min.

【关键词】 TTC脱氢酶活性标准曲线
【Key words】 TTCdehydrogenase activitystandard curve
【基金】 天津市自然科学基金(08JCYBJC13200)
  • 【文献出处】 实验室科学 ,Laboratory Science , 编辑部邮箱 ,2009年04期
  • 【分类号】X703
  • 【被引频次】57
  • 【下载频次】1377
节点文献中: 

本文链接的文献网络图示:

本文的引文网络