节点文献

New bienzymatic strategy for glucose determination by immobilized-gold nanoparticle-enhanced chemiluminescence

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【Author】 LIN JieHua,ZHANG Hui & ZHANG ShuSheng Key Laboratory of Eco-chemical Engineering,Ministry of Education,College of Chemistry and Molecular Engineering,Qingdao University of Science and Technology,Qingdao 266042,China

【摘要】 Bienzymatic biosensor for the determination of glucose by flow injection chemiluminescence(CL) de-tection was proposed.Hybrids of gold nanoparticles(GNPs) and chitosan were chosen as the immobi-lization matrix of glucose oxidase(GOD) and horseradish peroxidase(HRP) to fabricate the biosensors with silane-pretreated glass microbeads.After the enzyme catalyzing oxidation of glucose in GOD biosensor,the produced H2O2 flowed into HRP biosensor to react with luminol.The doped GNPs in chitosan were found to enhance the classical CL reaction of luminol-H2O2-HRP.The CL enhancement was investigated in detail by CL and UV-visible spectrum.Under the optimized experimental conditions,glucose could be determined in a linear range from 0.01 to 6.0 mmol/L with a detection limit of 5.0 μmol/L at 3σ.The accuracy of the proposed method was examined by detecting the glucose level in four clinical serum samples from hospital.The proposed method provides a new alternative to deter-mine glucose.

【Abstract】 Bienzymatic biosensor for the determination of glucose by flow injection chemiluminescence(CL) de-tection was proposed.Hybrids of gold nanoparticles(GNPs) and chitosan were chosen as the immobi-lization matrix of glucose oxidase(GOD) and horseradish peroxidase(HRP) to fabricate the biosensors with silane-pretreated glass microbeads.After the enzyme catalyzing oxidation of glucose in GOD biosensor,the produced H2O2 flowed into HRP biosensor to react with luminol.The doped GNPs in chitosan were found to enhance the classical CL reaction of luminol-H2O2-HRP.The CL enhancement was investigated in detail by CL and UV-visible spectrum.Under the optimized experimental conditions,glucose could be determined in a linear range from 0.01 to 6.0 mmol/L with a detection limit of 5.0 μmol/L at 3σ.The accuracy of the proposed method was examined by detecting the glucose level in four clinical serum samples from hospital.The proposed method provides a new alternative to deter-mine glucose.

【基金】 Supported by the Natural Science Foundation of Shandong Province (Grant No. Q2007B03);the Doctoral Fund of Qingdao University of Science and Technology (Grant No. 0022141);the National Natural Science Foundation of China (Grant No. 20775038)
  • 【文献出处】 Science in China(Series B:Chemistry) ,中国科学(B辑:化学)(英文版) , 编辑部邮箱 ,2009年02期
  • 【分类号】R446.1
  • 【被引频次】7
  • 【下载频次】74
节点文献中: 

本文链接的文献网络图示:

本文的引文网络