MR LAVA技术在腹部血管成像中的应用

Application of MR LAVA technique in abdominal vessels

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 吴何嘉史长征蔡香然李启权陈汉芳罗良平

【Author】 WU He-jia, SHI Chang-zheng, CAI Xiang-ran, LI Qi-quan, CHEN Han-fang, LUO Liang-ping (Medical Imaging Center, the first Affiliated Hospital of Jinan University, Guangzhou 510630, China)

【机构】 暨南大学附属第一医院医学影像中心暨南大学附属第一医院医学影像中心 广东广州510630广东广州510630

【摘要】 目的评价磁共振LAVA技术对上中腹部血管成像的应用价值与图像质量。方法44例用LAVA技术行上中腹检查并作动态增强的病例,按动脉血管树枝状走行分为五级,分析此方法对各级正常、变异及病变血管的显示能力,并与50例行3D DCE-MRA检查病例比较它们的血管成像质量。结果LAVA动态增强与3D DCE-MRA两种成像方法对Ⅱ、Ⅲ、Ⅳ、Ⅴ级血管的显示率差异均有显著统计学意义(P=0.000),LAVA技术中Ⅰ、Ⅱ、Ⅲ级组与Ⅳ、Ⅴ级组间两两比较有显著统计学意义(P=0.000)。LAVA动态增强与3D DCE-MRA这两种方法在血管成像质量方面没有统计学意义(χ2=0.660,P=0.719)。LAVA组病例中发现7例血管变异(其中6例动脉,1例静脉),1例病变血管。3D DCE-MRA对照组发现2例病变血管,由于范围所限无法评价变异血管。结论LAVA动态增强技术对正常血管显示率及血管变异情况明显优于3D DCE-MRA,加上LAVA具有较高的分辨率,可同时显示腹部脏器的病变,故在上中腹部血管成像中LAVA动态增强可代替3D DCE-MRA。

【Abstract】 Objective To evaluate the application value and image quality of the magnetic resonance LAVA technique for the vessels in the abdomen. Methods Forty-four cases examined with MR LAVA technique were collected, in which the abdominal arteries were classified into 5 grades according to their branch-like courses. Then the ability of LAVA for displaying normal, variant and pathological vessels was analyzed, and their vessels imaging quality was compared with 3D DCE-MRA in another 50 cases. Results The displaying rate of the vessels of Ⅱ, Ⅲ, Ⅳand Ⅴgrading showed a statistically evidence of statistical difference (P=0.000) between LAVA group and 3D DCE-MRA group. In LAVA group, a strong statistical significance was found when comparing vessels of Ⅰ,Ⅱ, Ⅲ grades with vessels of Ⅳ and Ⅴgrades (P=0.000). There was no statistical significance in the imaging qualily of the vessels between LAVA and 3D DCE-MRA (χ2=0.660, P=0.719). Seven cases with vascular variant (6 in artery and 1 in vein) and 1 case with pathological vessel were found in the LAVA group. In 3D DCE-MRA group only 2 cases with pathological vessels were found, and because of the limited scanning range vascular variant could not be evaluated. Conclusion LAVA technique is evidently superior to 3D DCE-MRA in displaying both normal vessels and vascular variant. In addition, LAVA has a higher resolution and can display abdominal foci simultaneously. Therefore, LAVA can substitute 3D DCE-MRA in abdominal vascular imaging.

【关键词】 磁共振成像LAVA技术3D DCE-MRA
【Key words】 Magnetic resonance imagingLAVA technique3D DCE-MRA