节点文献

数字X线机常规输出稳定性的测试

Stability measurement of digital X-ray unit on their daily output

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 孙峰张梦龙朱海涛

【Author】 SUN Feng,ZHANG Meng-long,ZHU Hai-tao Shandong Medical Imaging Research Institute,Jinan 250021,P.R.China

【机构】 山东省医学影像学研究所山东省医学影像学研究所 山东济南250021山东济南250021

【摘要】 目的:通过测试数字成像设备的X线机管电压和曝光时间稳定性,探讨数字成像摄影条件的选择原则。方法:在证明四台数字X线机mAs稳定性合格的前提下,选择合适的kVp档和时间档,测量其高压不准确度和曝光时间不准确度,并同国家标准进行比较。结果:四种X线机管电压不准确度在各个kVp档上均小于5%;时间不准确度大约在20ms档以上小于5%,在20ms档以下大于5%。结论:应用数字X线机摄影时曝光时间不可选择太小。

【Abstract】 Objective:To obtain suitable digital radiologic exposure factors of CR and DR by measuring stability of their X-ray tube voltage and exposure time.Methods:Four X-ray units were studied,the actual X-ray tube voltage was orderly measured when pre-setting voltage was 45,60,80,100,120,the actual exposure time was orderly measured when pre-set exposure time was 2,6,40,50,100,500,1000.By comparing actual measured values and pre-set ones,inaccuracy was obtained.Results:For all of four X-ray units,inaccuracy of voltage were less than 5%;inaccuracy of exposure time were less than 5% when pre-set time >20ms,and more than 5% when pre-set time < 20ms.Conclusion:In practice exposure time should keep high level to ensure X-ray unit output steady.

  • 【文献出处】 医学影像学杂志 ,Journal of Medical Imaging , 编辑部邮箱 ,2008年04期
  • 【分类号】R812
  • 【下载频次】35
节点文献中: 

本文链接的文献网络图示:

本文的引文网络