节点文献
微柱富集-石墨炉原子吸收光谱法测定痕量钼
Micro Column Preconcentration-GFAAS Determination of Trace Amounts of Molybdenum
【摘要】 提出和制备了一种负载苯基荧光酮(PF)的滤纸纤维柱,用于石墨炉原子吸收光谱法测定痕量钼的预富集。试验表明:在pH2.0时,控制待测液流速在2.0mL·min-1以内,可定量富集钼(Ⅵ),饱和吸附量为25.2μg·g-1。用流速1.0mL·min-1的氨水(3+97)2.0mL,可将5.0μg钼(Ⅵ)从吸附柱上定量洗脱。对国家地质标准物质GSS5和GSR1的分析结果表明,该方法可靠。测定了人工合成锌样、铜精矿中痕量钼,方法的检出限为1.01μg·L-1,回收率为92.0%~103.0%。
【Abstract】 A micro column packed with phenylfluorone coated fiber of filter paper was prepared.Mo(Ⅳ)can be quantitatively adsorbed on the micro column at pH 2.0 with flow rate of 2.0 mL·min-1and eluted from the column with 2.0 mL of aqueous ammonia(3+97)at flow rate of 1.0 mL·min-1.Amount of saturated adsorption of Mo(Ⅵ)on the column was 25.2μg·g-1.Accuracy of the method has been verified via the analysis of the standard reference materials GSS 5 and GSR 1.Detection limit of the method was 1.01μg·L-1and recovery was in range of 92.0%-103.0%.The method has been applied to the determination of traces of molybdenum in simulated samples of Zn and copper concentrates.
【Key words】 Molybdenum(Ⅵ); Micro column preconcentration; Preconcentration; GFAAS; Phenylfluoreone;
- 【文献出处】 理化检验(化学分册) ,Physical Testing and Chemical Analysis(Part B:Chemical Analysis) , 编辑部邮箱 ,2008年09期
- 【分类号】O614.612
- 【被引频次】5