节点文献

RP-HPLC二极管阵列检测器对阿替洛尔注射液质量的研究

RP-HPLC diode array detector for determinating of related substances of atenolol and its injection

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 刘玉真聂新永谢英花戎欣玉

【Author】 LTU Yu-zhen~* NIE Xing-yong XIE ying-hua RONG Xin-yu (Hebei University of Science and Technique College of Chemiscal and Pharmersutical,Shijiazhuang 1350018,China)

【机构】 河北科技大学化学与制药工程学院

【摘要】 目的:建立阿替洛尔注射液含量及有关物质的 RP-HPLC 测定法。方法:采用 Shimadzu ODS 色谱柱,以甲醇-磷酸二氢钾-庚烷磺酸钠(400:600:1.3)为流动相,检测波长275 nm。结果:阿替洛尔峰与各成分的分离度及检出灵敏度满足测定要求。阿替洛尔峰最小检出量为2ng(S/N=3)。二极管阵列检测器检测阿替洛尔峰纯度>99.99%。结论:本法简便,专属,准确。

【Abstract】 Objective:To establish a method for the determination of related substances in Atenolol and its injection.Methods: Shimadzu ODS column was used.The mobile phase was methanol:monopotassium phosphate:sodium heptane sulfonate(400:600: 1.30).Detection wavelength was at 254 nm.Results:Resolution and detectable sensitivity of each constituentin sample satisfies determined demand of related substances.The limit of detection was 2 ng(S/N=3).The peakpurity of Atenolol determined by diode array detector is >99.99%.Conclusion:This method is found to be simple,selective,effective.

  • 【文献出处】 药物分析杂志 ,Chinese Journal of Pharmaceutical Analysis , 编辑部邮箱 ,2006年03期
  • 【分类号】R927
  • 【被引频次】3
  • 【下载频次】71
节点文献中: 

本文链接的文献网络图示:

本文的引文网络