人工颈椎间盘置换术后颈椎活动与稳定性的评价

Evaluation of cervical spine activity and stability with Bryan cervical prosthesis replacement

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 梁秋发原林王文军曹盛俊刘利乐朱一平王麓山姚女兆全必春

【Author】 LIANG Qiu-fa,YUAN Lin,WANG Wen-jun,et al. Key Laboratory of Medical Biomechanics of Southern Medical University,Guangzhou 510515,China

【机构】 南方医科大学生物力学重点实验室南华大学附属第一医院脊柱外科南华大学附属第一医院脊柱外科 广州510515广州510515湖南衡阳421001

【摘要】 目的:评价人工颈椎间盘置换术对颈椎活动与维持手术节段稳定性的作用及临床效果。方法:18例非手术治疗无效的颈椎病(脊髓型、神经根型或/和混合型)行前路颈椎间盘切除后置入Bryan人工颈椎间盘,按计划进行定期随访(6~20个月,平均14个月),根据手术前后颈椎的动力位X线片分析手术节段的稳定性。并结合病人神经症状和体征的改善情况评价其临床效果。结果:手术效果优9例,良5例,一般3例,差1例。手术前平均JOA评分为12分,术后平均为15.5分;手术节段前屈后伸术前运动范围12.8°±5.7°,术后11.2°±5.4°;邻近间隙术前屈伸运动范围11.9°±5.8°,术后11.4°±4.9°。术前后比较无统计学差异(P>0.05)。未发现假体松动和移位。结论:人工颈椎间盘置换术安全可靠,病人恢复快,既可保持手术节段颈椎稳定,又可维持颈椎的生理活动度。

【Abstract】 Objective: To determine whether the Bryan disc prosthesis can provide relief and maintain stability and segmental motion or not through objective neurologic symptoms and signs evaluation. Methods: 18 patients(22 discs) with symptomatic cervical radiculopathy andor myelopathy underwent implantation with the Bryan prosthesis after a standard anterior cervical discectomy. As scheduled follow-up periods(6 to 20 months, average 14 months), the effectiveness of the device was characterized by evaluating each patient’s pain, neurologic function, and radiographically measured range of motion at the implanted level. Results: JOA improved from average 12 grades preoperatively to 15.5 postoperatively. The dynamic radiographs showed that the flexion-extension range of motion per level averaged from preoperative 11.8±5.8 degrees to postoperative 11.2±5.4 degrees, and from 11.9±5.8 degrees to 11.4±4.9 degrees at the adjacent segment(P>0.05). No devices explanted. Conclusions: Discectomy and implantation of the device alleviates neurologic symptoms and signs similar to anterior cervical discectomy and fusion. Radiographic evidence supports maintenance of motion. The procedure is safe and the patients recover quickly. Successive follow-up will be needed to assess the long-term functionality of the prosthesis and protective influence on adjacent levels.

  • 【文献出处】 中国临床解剖学杂志 ,Chinese Journal of Clinical Anatomy , 编辑部邮箱 ,2005年06期
  • 【分类号】R687.3
  • 【被引频次】10
  • 【下载频次】215
节点文献中: 

本文链接的文献网络图示:

本文的引文网络