节点文献
动态流导法真空校准装置
Vacuum calibration system by means of dynamic flow conductance
【摘要】 动态流导法真空校准装置是真空计量的标准,可用于高真空和超高真空规的校准,真空度的校准范围为10-1~5×10-7Pa,合成标准不确定度为0.69%~1.4%。
【Abstract】 Dynamic flow conductance is used to calibrate vacuum systems to offer a metrological standard for vacuum, which is available to the calibraction of HV and UHV gauge. Its vacuum calibration range is from 10-1 to 5×10-7 Pa, with an uncertainty of the standard for combined calibration up to 0.69%~1.4%.
- 【文献出处】 真空 ,Vacuum , 编辑部邮箱 ,2005年03期
- 【分类号】TB771
- 【被引频次】15
- 【下载频次】217