节点文献
电动血压计与汞柱式血压计测压的比较
The comparison of function between electronic sphygmomanometer and mercury column type sphygmomanometer.
【摘要】 目的:评价电子血压计测量血压的准确性。方法: 200名体检者用BP-203RVⅡ型电动血压计同时与台式汞柱血压计测量血压,进行自身对照并比较两种测量结果。结果:两种血压测量结果差异无显著性 (P>0.05)。结论:BP-203RVⅡ型电动血压计测定血压可靠、方便、客观,有良好的应用前景。
【Abstract】 Objective: To appraise the accuracy of electronic sphygmomanometer in measuring the blood pressure. Method: 200 physical examinee were measured blood pressure with the desk-top mercury column sphygmomanometer at the same time with the BP-203RVⅡ type electronic sphygmomanometer, carried on one’s own contrast and compare the two kinds of measurement results. Result: There was no significant difference between the two kinds of results. Conclusion: BP-203RVⅡ type electronic sphygmomanometer is reliable, convenient, objective to determine blood pressure, and there is good application prospects.
【Key words】 the blood pressure determining; sphygmomanometer; randomized contrast experiment;
- 【文献出处】 齐鲁护理杂志 ,Journal of Qilu Nursing , 编辑部邮箱 ,2005年01期
- 【分类号】R472.9
- 【被引频次】8
- 【下载频次】139