氧化电位水对消化内镜消毒效果的研究

The effectiveness of electrolyzed oxidizing water in disinfections of endoscopes

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 李建辉周丽雅林三仁杨雪龄丁士刚董秀云

【Author】 LI Jian-hui, ZHOU Li-ya, LIN San-ren, et al. Department of Digestion, Third Hospital, Peking University, Beijing 100083, China

【机构】 北京大学第三医院消化科北京大学第三医院消化科北京大学第三医院消化科 100083河北省承德市中心医院消化科 067000100083

【摘要】 目的 检测氧化电位水(electrolyzed oxidizing water,EOW)对消化内镜的消毒效果并与2%戊二醛相比较。方法 (1)对一般细菌的消毒效果:顺次抽取检查患者后的消化内镜,EOW组和2%戊二醛组均分别浸泡消毒1、3、5 min,每时间组均为30例,进行细菌培养。(2)对幽门螺杆菌(Hp)的消毒效果:顺次抽取快速尿素酶试验阳性者使用过的以及Hp人为污染后的内镜,2种消毒剂按消毒时间分组同上,PCR法检测Hp残留。(3)对乙肝病毒(HBV)的消毒效果:用HBV阳性血清稀释液污染内镜,2种消毒剂分别作用1、3、5、10 min,每时间组30例,PCR法检测HBV残留。结果(1)一般细菌:消毒1、3、5 min后,EOW组细菌培养均为阴性,2%戊二醛组分别有阳性5例、4例、1例。(2)Hp:EOW组作用1 min后2例阳性,3、5 min后的均为阴性;2%戊二醛作用1、3、5 min后分别有3例、4例、4例阳性。(3)HBV:EOW组作用1、3、5、10 min各组均为阴性;2%戊二醛组作用后各时间组分别有2例、5例、4例、5例阳性。结论 EOW对一般细菌、Hp、HBV等有较强的杀灭作用,较2%戊二醛消毒力强、快速,可推荐为临床内镜消毒。

【Abstract】 Objective To observe the effectiveness of electrolyzed oxidizing water(EOW) in disinfections of endoscopes and compared with that of 2% glutaraldehyde. Methods (1) Disinfections of bacteri-a:Specimens were obtained immediately from endoscopes after being used, disinfected and washed out for bacterial culture. Endoscopes were immersed in either EOW or 2% glutaraldehyde for 1 , 3 and 5 minutes. Each time group consisted of 30 cases. (2 ) Effectiveness on disinfections of H. Pylori: endoscopes either from patients with H. Pylori infection or artificially contaminated with H. Pylori were washed out and exposed for 1 ,3 and 5 minutes in each of the disinfectants. PCR was used to detect DNA of H. Pylor. (3) Effectiveness of disinfections of HBV : Endoscopes were contaminated with HBV-positive serum, and the exposure time to both disinfectants were 1,3,5 and 10 minutes respectively. Thirty samples were studied in each time groups. Results ( 1 ) After manual cleaning and disinfections for 1, 3 and 5min in EOW, no vegetative bacteria were found. In 2% glutaraldehyde processing group, 5/30, 4/30 and 1/30 specimens remained bacteria positive, respectively. (2)There was no positive specimens found after 3 and 5 minutes’ exposure in EOW in the group of H. Pylori, but 2/30 positive specimens were found in the 1 minute group. While 3/30, 4/30 and 4/30 positive specimens were found after exposure in 2% glutaraldehyde. (3) In the group of HBV, there is no positive case found after exposure in EOW, while in 2% glutaraldehyde groups, there were 2,5 ,4 and 5 cases remained positive in 1 ,3,5 and 10 minutes group respectively. Conclusion EOW is superior to 2% glutaraldehyde in disinfections of Hp or HVB with high-speed and thoroughness, and worthy to be recommended in disinfections of endoscopes.

  • 【文献出处】 中华消化内镜杂志 ,Chinese Journal of Digestive Endoscopy , 编辑部邮箱 ,2004年03期
  • 【分类号】R187
  • 【被引频次】19
  • 【下载频次】93
节点文献中: 

本文链接的文献网络图示:

本文的引文网络