节点文献

深圳迈瑞BS-300全自动生化分析仪性能评价

Performance evaluation of BS-300 automatic biochemistry analyzer

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 莫其农周小梅何进才李伟雄梁广佳

【Author】 MO Qinong, ZHOU Xiaomei, HE Jincai, LI Weixiong, LIANG Guangjia. Department of Clinical Laboratory, Shenzhen Hospital of Traditional Chinese Medicine, Guangdong Shenzhen 510083, China

【机构】 深圳市中医院检验科深圳市中医院检验科 广东深圳518033广东深圳518033广东深圳518033

【摘要】 目的 评价BS 30 0全自动生化分析仪主要性能。方法 选定涵盖各波长的 9个项目 [丙氨酸氨基转移酶 (ALT) ,γ 谷氨酰基转移酶 (GGT) ,葡萄糖 (Glu) ,总蛋白 (TP) ,白蛋白 (Alb) ,肌酐 (Cr) ,钙 (Ca) ,氯 (Cl-) ,载脂蛋白B(ApoB) ],对深圳迈瑞BS 30 0全自动生化分析仪进行精密度、线性、交叉污染和对比试验 ,并对医学决定水平浓度进行偏倚估计。结果 BS 30 0日间、批内、批间及总CV均 <4 % ;线性稀释变异P <P0 .0 5(0 .6 84 1 ) ,线性失拟G <F0 .0 5(3.2 9) ;交叉污染率 2 .81E - 0 6 ;与AU2 70 0相关系数 (r) >0 .96 ;在医学决定水平浓度处的相对偏倚ALT、GGT、Glu、TP、Alb、Cr、Ca、Cl-(90mmol/L)、Cl-(1 2 0mmol/L)、ApoB分别为- 1 .8%、- 0 .1 %、- 3.9%、- 0 .5 %、- 2 .3%、- 7.2 %、0 .4 %、3.9%、3.2 %、- 1 .7%。结论 BS 30 0全自动生化分析仪主要性能指标达到了NCCLS标准 ,接近或达到国外同类产品要求 ,结果可靠 ,适合中小型医院使用

【Abstract】 Objective To evaluate the main performance of BS 300 automatic biochemistry analyzer made in a Shenzhen Mindray Co. Methods To evaluate the precision, linearity crossing contamination and contrasting test of BS 300 automatic biochemistry analyzer by 9 items that contain all wavelength (ALT, GGT, Glu, TP, Alb, Cr, Ca, Cl -, ApoB). It was estimated its bias at medicine determined level concentration. Results The daytime run CV , within run CV , between run CV and total CV were under 4%. The diluted variance of linearity P was less than P 0.05 (0.684 1). The linear drift G was less than F 0.05 (3.29). The cross contamination rate was 2.81E 06, and correlated factor r with AU2700 was greater than 0.96. The relative biases at determined medicine level concentration with ALT,GGT,Glu,TP,Alb,Cr,Ca,Cl -(90 mmol/L), Cl -(120 mmol/L),ApoB were -1.8%,-0.1%, -3.9% ,-0.5%,-2.3%,-7.2%,0.4%,3.9%,3.2%,-1.7%. Conclusions The main performance of BS 300 automatic biochemistry analyzer accords with NCCLS, and is comparable to imported instrument. It is appropriate for small or middle sized hospital.

  • 【文献出处】 检验医学 ,Shanghai Journal of Medical Laboratory Sciences , 编辑部邮箱 ,2004年04期
  • 【分类号】TH789
  • 【被引频次】6
  • 【下载频次】176
节点文献中: 

本文链接的文献网络图示:

本文的引文网络