采用移动极差统计合并技术来监控ICP-AES测量系统的不确定度
Uncertainty in-Statistical-Monitor for ICP-AES Measurement System by Using Moving Range Pooled
【摘要】 采用ICP AES(电感耦合等离子体 原子发射光谱法 )测定金属硅中铁杂质含量进行不确定度评估时 ,提出了平均移动极差 (MR)合并统计动态监控技术。研究认为 ,采用该技术来进行ICP AES测量系统的不确定度评估 ,其贡献量分别是由控制样品 (QC)和线性拟合两部分合成。这一结论的得出能最大限度合成各种变异因素的影响 ,避免了相关性的复杂计算 ,有利于ICP AES测量系统的不确定度评估
【Abstract】 This paper describes how to estimate the uncertainty when Fe in metallic silicon is determined by ICP-AES (Inductively Coupled Plasma Atomic Emission Spectrometry). The technique of the averaged moving range((MR)) pooled in-statistical-dynamic-monitor is proposed. The study suggests that the uncertainty contribution of ICP-AES measurement system is, by using this technique, a combination of quality control (QC) and linear fit. This conclusion maximized the combination of the effects on various variances, (refrained) from the complicated relativity, and favored the uncertainty evaluation for ICP-AES measurement system.
【Key words】 ICP-AES; Metallic silicon; Uncertainty; (MR) pooled in-statistical-monitor; Quality control; Linear fit; AD normality (statistics); Autocorrelation trend analysis; Moving range; Exponentially weighted moving average; Horwitz equation;