X射线平板探测器数字成像及其图像校准
X-ray flat-panel-detector-based digital radiography and its image calibration
【摘要】 为了提高X射线数字成像系统的图像质量 ,以基于平板探测器的X射线数字照相系统为研究对象 ,介绍了平板探测器自身影响图像质量的 4大因素 :随机噪声 ,偏置误差 ,像元响应不一致性和瑕疵像元等 ,分析了这些因素的成因及其对图像质量的影响 ,提出了相应的减小和消除这些影响的图像校准方法 .实验结果表明 :运用这些图像校准方法 ,可以有效地去除上述因素对图像质量的影响 ,显著提高平板探测器输出图像的质量 ,为后续图像判读奠定了良好的基础
【Abstract】 To improve imaging quality of X-ray digital radiography system, an X-ray flat-panel-detector-based digital radiography system was investigated. Four factors of flat panel detector which degrade image quality, such as random noise, offset, response nonuniformity and defective pixels, were introduced. How they occur and influence imaging quality was analyzed. Corresponding methods for image calibration were put forward to minish and eliminate the influences on image quality. Experimental results indicate that, with the image calibration methods, the influences of above mentioned factors can be removed effectively and the quality of image from flat panel detector can be improved greatly, which has founded a good foundation for succedent image identifying.
【Key words】 non-destructive tests; image processing; image calibration; flat panel detector;