节点文献

中山市医用诊断X射线机质量调查分析

Analysis on data of image quality control tests for medical X-ray diagnosis machines in Zhongshan city

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 关康年

【Author】 GUAN Kang nian (Zhongshan Center for Diseases Control and Prevention, Zhongshan 528403, China)

【机构】 中山市疾病预防控制中心 广东中山528403

【摘要】 目的 了解中山市医用诊断X射线设备的运行性能状况。方法 使用瑞典Solidose 40 0数字化X射线机影像质量控制检测仪 ,对 16项指标进行质控监测。结果 检测了 67台X射线诊断机 ,检测指标项次为 765项 ,有 45 3项次合格 ,合格率为 5 9 2 % ;16项指标全部合格的仅有 3台 ,占 4 5 %。结论 中山市医用诊断X射线机合格率偏低 ,应严把质量控制关 ,提高影像质量水平

【Abstract】 Objective To explore the working performance status of medical X ray diagnosis machines. Methods 16 indexes of image quality control for a X ray diagnosis set were measured using Solidose 400 digital X ray meters. Results 67 sets of X ray diagnosis machines and 765 items were measured, including 453 qualified items and the qualified rate was 59 2%; only 3 sets for 16 items were fully qualified and the rate was 4 5%. Conclusion It would be necessary to enhance image quality control test.

  • 【文献出处】 中国职业医学 ,Chinese Occupational Medicine , 编辑部邮箱 ,2003年02期
  • 【分类号】R197.39
  • 【被引频次】1
  • 【下载频次】22
节点文献中: 

本文链接的文献网络图示:

本文的引文网络