节点文献

差热分析法测定多元醇的比热

Heat Capacity of Glycol Determined by Differential Scanning Calorimeter

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 杨长生马沛生夏淑倩

【Author】 YANG Changsheng, MA Peisheng, XIA Shuqian(School of Chemical Engineering and Technology, Tianjin University, Tianjin 300072, China)

【机构】 天津大学化工学院天津大学化工学院 天津300072天津300072天津300072

【摘要】 利用差示扫描量热计测定了乙二醇、丙二醇、丁二醇和叔丁醇在不同温度范围内的比热.为了验证仪器的准确性,实验测定二次蒸馏水在20~80℃的比热,并与文献值进行了对比.测定水时仪器测量误差在2%以内,并且把测得的各种醇的比热与文献值进行了对比,比热测量值的精度完全能满足工程计算的要求.同时回归出了它们的Cp随温度的变化关系,可供工程计算中参考.

【Abstract】 Measurements of the heat capacities of the ethylene glycol, 1,2propanediol , 1,4butanediol and tertbutanol were performed at different temperature range with a differential scanning calorimeter(DSC). The heat capacity of liquid water from 20℃~80℃ has been measured to verify the validity of the Cp measurements. The measured values are compared with the literature data. The error is less than 2%. The measured Cp of each glycol has been expressed as a function of temperature. The measured heat capacities are of sufficient accuracy for most engineeringdesign calculation.

【基金】 中国石油化工股份公司科技开发基金资助项目(200049)
  • 【文献出处】 天津大学学报 ,Journal of Tianjin University , 编辑部邮箱 ,2003年02期
  • 【分类号】O657.99
  • 【被引频次】7
  • 【下载频次】403
节点文献中: 

本文链接的文献网络图示:

本文的引文网络