PSPO-CCD系统评估
THE EVALUATION OF PSPO-CCD SYSTEM
【摘要】 本文介绍对分光仪PSP0探测器TH7832 CCD的评估结果。该CCD系统工作在-130℃~-110℃温度下,具有较好的稳定性和很好的线性。系统的读出噪声约为300电子,动态范围3500。在实验结果的基础上,文章对系统性能的改善进行了简单的讨论。
【Abstract】 The evaluation results of the CCD system in Prism-Spectro Photometer, PSPO, are described. The instrument employs TH7832 CCD as detector, which works at temperatureof - 130℃--11 0℃ . The system has good stability and linearity. Readout noise is about300 electrons and dynamic range 3500. The quantum efficiency of the detector will be calibrated with the optical system in real observational environment. Further improvements on the readout circuit and system gain are discussed.
- 【文献出处】 紫金山天文台台刊 , 编辑部邮箱 ,1998年03期
- 【分类号】TH751
- 【下载频次】24