东芝TBA30—FR全自动生化分析仪评价

Evaluation of TBA-30FR Automatic Biochemical Analyzer of TOSHIBA

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 刘晨梅; 王兆成; 冯蔼苓; 管卫;

【Author】 Liu Chenmei; Wang Zhaocheng ;Feng Ailing; Guan Wei (Tianjin Childern’s hospital Biochemical laboratory, Tianjin 300074)

【机构】 天津市儿童医院检验科!天津; 300074;

【摘要】 评价日本东芝TBA-30FR全自动生化分析仪的工作性能。方法:用进口和国产试剂及定值质控血清对该仪器的准确度、精密度及交叉污染等进行了观察。结果:应用进口及国产试剂测定生化指标20项,其准确度除1个结果位于允差范围内外,其余结果均位于可信范围内。精密度观察:批内不精密度cv%为0.27~0.71,批间不精密度cv%为0.87~2.88。交叉污染率为0~0.87%。结论:通过准确度、精明度和交叉污染试验的观察,我们认为该仪器可以满足临床化学实验室的要求。

【Abstract】 Obective: To evaluate the capability of TBA3oFR automatic biochemical analyzer. Method: To determine the accurary, precision and carry over experiment of the instrument with the reagent of forgien and national preduct. Results: The accurary of 20 tests in the instrument were in target range except for 1 result, which was in the recommended value. The imprecision of intra-and inter-day cv% were 0. 27-0. 71 and 0. 87-2. 88 %. The carry over ratio was 0-0. 87%. Conclusion: The results showed, that the instrument can be used for clinical chemistry laboratory.

  • 【文献出处】 现代科学仪器 ,MODERN SCIENTIFIC INSTRUMENTS , 编辑部邮箱 ,1998年06期
  • 【分类号】TH77
  • 【被引频次】2
  • 【下载频次】27