节点文献
在线氧化-流动注射-化学发光法测定痕量Cr(Ⅲ)/Cr(Ⅵ)
ON-LINE OXIDATION FLOW-INJECTION CHEMILUMINESCENCE METHOD FOR DETERMINATION OF TRACE AMOUNTS OF Cr(Ⅲ) AND Cr(Ⅵ)
【摘要】 研究了在线氧化-流动注射-化学发光法测定痕量Cr(Ⅵ)和Cr(Ⅲ)的方法,基于槲皮素-过氧化氢-Cr(Ⅵ)-氢氧化钾化学发光直接测定Cr(Ⅵ),Cr(Ⅲ)经二氧化铅柱在线氧化后测定。方法简单、快速,Ct(Ⅵ)检出限为1.0×10-9g·ml-1,Cr(Ⅵ)线性范围为1.0×10-8~1.0×10-6g·ml-1,用于环境水样分析,结果满意。
【Abstract】 An on-line oxidation flow-injection chemiluminescence method for determination of Cr( Ⅱ ) and Cr( Ⅵ ) is established and reported in this paper. Hexavalent chromium is directly measured with Quercetin -H2O2-KOH chemiluminescence reaction and the determination of Cr ( Ⅱ ) is carried out after on-line oxidation of Cr( Ⅲ ) to Cr( Ⅵ ) with PbO2 column at 85℃. The detection limit is 1. OX10-9g ?ml-1 for Cr(Ⅵ ) with a linear range from 1. OX10-81.0X10-6g ?ml-1. The proposed method has been applied to the analysis of Cr( Ⅵ ) and Cr( Ⅲ ) in environmental water samples with satisfactory results.
【Key words】 On-line oxidation; Flow-injection; Chemiluminescence; Valence states analysis; Chromium;
- 【文献出处】 理化检验(化学分册) ,Physical Testing and Chemical Analysis Partb Chemical Analgsis , 编辑部邮箱 ,1998年07期
- 【分类号】O657.3
- 【被引频次】21
- 【下载频次】99