节点文献
用于微剂量学研究的无壁正比计数器系统
Using a Cylindrical Wall less Proportional Counter System for Studying Microdosimetry
【摘要】 本文介绍了为模拟荷能重离子在等效生物组织中径迹和能量沉积分布研究而设计、制造的无壁正比计数器和探测系统,并利用放射源239Pu的α粒子(能量为5.155MeV)在500Pa气压及工作气体为CH4(10%)+Ar(90%)下,对本探测器系统的分辨率与所加高压的关系进行了实验,并就利用此探测器系统研究微剂量学的研究方法进行了讨论。
【Abstract】 The structure and character of a cylindrical wall less proportional counter and its measurement system are introduced in the paper, which was designed to simulate the track structure and energy deposition distribution of heavy ion in tissue equivalent gas. In the experiment, 239 Pu α source was used to study the variation of FWHM of the detector system with the anode voltage at the pressure 500 Pa. A research method of microdosimetry was discussed by using the detection system. (
【关键词】 无壁正比计数器;
组织等效气体;
能量沉积;
离子径迹结构;
【Key words】 Wall less proportional counter Tissue equivalent gas Energy deposition Ion track structure);
【Key words】 Wall less proportional counter Tissue equivalent gas Energy deposition Ion track structure);
【基金】 中科院“八五”重点课题
- 【文献出处】 核电子学与探测技术 ,NUCLEAR ELECTRONICS & DETECTION TECHNOLOGY , 编辑部邮箱 ,1998年02期
- 【分类号】TL811.2
- 【被引频次】2
- 【下载频次】80