埋藏式心脏复律除颤器与单极导线起搏器的合并应用一例报告

Combined use of implantable cardioverterdefibfillator with unipolar pacemaker

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 任自文丁燕生陈健吴林张筠华

【Author】 Ren Ziwen, Din Yansheng, Chen Jian, et al. Department of Cardiology, First Hospital of Beijing Medical University, Beijing 100034

【机构】 北京医科大学第一医院心内科

【摘要】 国内首次合并应用埋藏式心脏复律除颤器(ICD)与单极起搏器1例。术中仔细观察起搏器对ICD的影响,未发现ICD对起搏心律的双感知,亦未发现对心室颤动(室颤)的感知不足,ICD对3次诱发的室颤均迅速感知并一次除颤成功。术后189天储存资料显示患者共自发21次室性心动过速,均被一次抗心动过速起搏有效终止,未发生误放电,亦未漏诊快速心律失常,初步观察效果满意。

【Abstract】 This is the first case combining use of implantable cardioverterdefibrillator (ICD) with unipolar pacemaker in our country. The effect of pacemaker on ICD was carefully evaluated during implantation. There were no double sensing of ICD to pacing rhythm and no undersensing to ventricular fibrillation (VF). ICD rapidly detected the induced VFs in three tests and successfully defibrillated the Vf by one shock in every test. The stored data of 189 days postimplantation showed that total twenty one episodes of spontaneous ventricular tachycardias were detected and every episode was successfully terminated by only one burst of antitachycardiac pacing. There was no faulty delivery of therapy and no loss of detecting tachyarrhythmias. The result of combined ICD with unipolar pacemaker was satisfying in the preliminary observation.

  • 【文献出处】 中华心血管病杂志 ,CHINESE JOURNAL OF CARDIOLOGY , 编辑部邮箱 ,1997年06期