节点文献
轻烃中痕量硫的紫外荧光分析系统的改进
The Development of Analytical System and Method for the Determination of Trace Sulfur in Light Hydrocarbons
【摘要】 开发了一种新的 0.1~1.0mg/L痕量硫的分析系统及方法。方法以 Antek 7000氧化裂解-紫外荧光定硫仪为基础,通过对进样系统、裂解管、及检测限等方面的改进,使单次分析样品最大进样量增加到 50μL。方法的检测限达到 0.1mg/L,满足了科研及生产工艺的需要。
【Abstract】 An improved system and method of 0. 1-1. 0 mg/L trace sulfur detection in light hydrocarbons was developed on the Antek 7000 Pyrolysis UV-induced Fluorescence Sulfur Analyzer. The maximum injection volume was raised to 50μL,after heating the pipeline con- necting the pyrotube and the detection system and using a new design pyrotube. The results showed that the method is good enough to meet the need of the technological research and in- dustrial process with a detection limit of 0. 1mg/L.
- 【文献出处】 分析化学 ,Chinese Journal of Analytieal Chemistry , 编辑部邮箱 ,1997年01期
- 【分类号】O657.3
- 【被引频次】1
- 【下载频次】73