节点文献
中子辐射损伤HPGe探测器的修复系统
System for Recovering Neutron Radiation Damaged HPGe Detectors
【摘要】 在束γ实验中,锗探测器因受中子辐射损伤而性能变差、能量分辨率变坏。本文介绍了自建的锗γ探测器中子辐射损伤修复系统的组成和工作情况。用该系统已成功地修复了4台N型HPGe探测器。
【Abstract】 Fast neutron produced in in-beam γ ray experiments can result in the energy resolutiondegradation of HPGe detectors. This paper intreduces a system set up in our laboratory for recov-ering neutron damaged germanium detectors and illustrates the recovering procedure. Using thissystem, four neutron damaged n-type HPGe detectors have been recovered successfully.(
【基金】 HIRFL国家实验室和国家自然科学基金
- 【文献出处】 核电子学与探测技术 ,NUCLEAR ELECTRONICS & DETECTION TECHNOLOGY , 编辑部邮箱 ,1996年03期
- 【分类号】TL816.2
- 【下载频次】64