节点文献

超声雾化空气-氩ICP-AES最佳工作条件的研究

Optimal cxperimental conditions of air-argon ICP-AES using ultrasonic nebulizer

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 张俊卿杨文斌

【Author】 Zhang Junqing (Departmentof Metallurgical Engineering,Baotou University of Iron and Steel Technology,Baotou,014010),Yang Wenbin (Department of Chemistry,Neimenggu University,Hohhot,010031)

【机构】 包头钢铁学院冶金系内蒙古大学化学系 包头014010呼和浩特010031

【摘要】 将医用超声雾化器改装成了电感耦合等离子体发射光谱仪(ICP—AES)的进样装置。用改装的超声雾化器进样,研究了多元素同时分析的最佳实验条件。在高频功率为3.8×0.80KW、载气流量0.80L/min和观察高度1.0cm的条件下,大多数元素谱线较强而背景辐射较弱。

【Abstract】 In order to improve the sample injection technique of ICP-AES, a medical ultrasonic nebulizer was modified. The optimal experimental conditions for simultaneous detennination of multielement were examined by using the modified nebulizer for sample injection. When the high-frequency induction power is 3. 8×0. 80kW ,the flow rate of carrier gas is 0. 80L/min and the observation height is 1.0cm,the intensity of spectral line is strong and the background is weak for most elements.

  • 【文献出处】 分析仪器 ,Analytical Instrumentation , 编辑部邮箱 ,1996年03期
  • 【分类号】TH744.3
  • 【被引频次】1
  • 【下载频次】43
节点文献中: 

本文链接的文献网络图示:

本文的引文网络