节点文献
流式细胞计多参数测量中的补偿问题
COMPENSATIVE PROBLEM IN MULTI-PARAMETER MEASUREMENT OF FLOW CYTOMETRY
【摘要】 目的:正确解决流式细胞计多参数测量的补偿问题。方法:应用外周血淋巴细胞的双阴性对照、双染样品电子补偿和二维点图得到正确的补偿值。结果:补偿值大小受PMT电压值和染色条件的影响。结论:流式测量中除设定实验样品的阴性对照外,在多重染色样品测量时还应设置阳性对照,以调整补偿器的正确值。
【Abstract】 Objective:To solve correctly the compensative problem in multi-parameterment measure-ment on flow cytometry.Methods:By using lymphocytes of human peripheral blood samples,double negtive control,electron compensator and dot plot,the correct compensation value was obtaind.Re-sults:The compensative value was influenced by PMT voltage and staining condition:of samples.Con-clusions:Except the negtive control set up in experiments,the positive controls are needed for multipa-rameter measurement in folw cvtometry.
- 【文献出处】 北京医科大学学报 ,JOURNAL OF BEIJING MEDICAL UNIVERSITY , 编辑部邮箱 ,1996年01期
- 【分类号】R446.113
- 【被引频次】1
- 【下载频次】83