离子交换预浓集-流动注射离子选择电极法测定水中氟
Determination of Fluoride in Water by Ion Exchange Preconcentration-Flow Injection Ion-selective Electrode
【摘要】 本文利用强碱型阴离子交换树脂对氟离子预浓集,与流动注射离子选择电极检测相结合建立了氟的高灵敏度、快速分析方法.柱子的富集效率达到50倍. 对于0.lμg/mL试样的相对标准偏差为3.2%(n=10),检出限为0.012μg/mL.利用通用型可调入射距离的喷壁型流通池,不但灵敏度高,而且可用于其它离子的测定.采用双柱交替采样,加快了分析的速度.
【Abstract】 The fluoride was concentrated by the strong base resins, then it was removed when the eluant flowed through the column, finally it was determined by ion-selective electrode. The sensitivity of fluoride is 50 times than that in flow injection analysis in which ion exchange column was not used, the relative standard deviation was 3. 2% for 0. 1 μg/mL fluoride (n=10). The detection limit is 0. 012μg/mL. A large-volume wall-jet cell was used for every ion-selective electrode. The distance between inlet nozzle and electrode surface can be changed through the position of inlet tube. The sampling rate was increased when double column was used.
【Key words】 Fluoride; ion exchange; ion-selective electrode; flow injection;
- 【文献出处】 分析化学 ,Chinese Journal of Analytieal Chemistry , 编辑部邮箱 ,1995年06期
- 【分类号】O657.1