节点文献
氢化物发生—原子吸收光谱法测定851口服液中的Se
Determination of Se in 851 Oral Liquid by Hydride Generation-AAS
【摘要】 本文研究了氢化物发生—原子吸收光谱法测定硒的适宜条件,考察了多种离子对硒测定的干扰情况,提出了消除某些过渡金属对硒测定干扰的方法。本方法用于851口服液测定时,对于含硒4μg/L 左右的样品,全程相对标准偏差<14%,加标回收率为93~108%。
【Abstract】 This paper discribed the determination of microelement Se in 851 oral liquid by Hy- dride Generation-AAS.The interferences of transition elements on the analysis of Se could by overcomed by adding Fe++ and EDTA.The precision of the method was<14% for the sample in which Se concentration was around 4μg/L.The recovery was 93~108%.
- 【文献出处】 福建分析测试 ,Fujian Analysis & Testing , 编辑部邮箱 ,1995年04期
- 【分类号】O657.31
- 【被引频次】1
- 【下载频次】18