钒(Ⅴ)-亚铁氰化钾-鲁米诺化学发光体系测定钒
Determination of Vanadium by FIA with V(Ⅴ)-Fe (CN)64--Luminol Chemiluminescence System
【摘要】 本文利用钒(Ⅴ)-亚铁氰化钾-鲁米诺化学发光体系建立了痕量钒的化学发光测定法。方法检出限是2×10-10g/mL钒,线性范围是4×10-10~1×10-7g/mL钒,测定的相对标准偏差小于2%,考察了26种常见离子的干扰情况。方法已用于水样中痕量钒的测定。
【Abstract】 In this paper,a new method for the determination of vanadium by FIA with V (V)-Fe (CN)64-- luminol chemiluminescence system has been reported. The detection limit is 2×10-10g/mL vanadium. The linear range of the caliberation curve is 4×10-10~1×10-7g/mL vanadium. The relative standard deviation is less than 2% (8 ×10-10g/mL vanadium,n=11). The method has been applied to determine vanadium in water samples.
【基金】 国家自然科学基金
- 【文献出处】 分析试验室 ,Analytical Laboratory , 编辑部邮箱 ,1992年01期
- 【被引频次】7
- 【下载频次】63