节点文献
直流磁化测量中样品长度修正因数的分析
ANALYSIS OF CORRECTION FACTOR FOR SAMPLES’ LENGTHS IN MEASUREMENT OF DC MAGNETIZATION
【摘要】 本文基于磁化电流在空间中磁场分布的计算,对不同长度的超导样品在测量线圈中真正排磁通作了分析,导出简单、明瞭的由检测线圈参数和样品长度所决定的修正因数。并将BiPbSrCaCuO高温单相样品截成不同长度,进行实验验证,实验结果与本文中的分析符合得较好。为小截面超导样品绝对抗磁矩的测量和计算提供了一种准确的分析方法。
【Abstract】 In this paper, we analysed the diamagnetization flux of samples with different lengths by calculating the distribution of the magnetic field produced by magnetization current. A simple and clear correction factor which was determined by the coil size and samples’ lengths was obtained. An accurate calculation method for absolute magnetization of samples with small cross section was presented. Through the measurement of the susceptibility on BiPbSrCaCuO superconductors with different lengths, this method is shown to be valid.
【基金】 国家超导技术联会研究开发中心;中国科学院超导办公室资助的课题
- 【文献出处】 物理学报 ,Acta Physica Sinica , 编辑部邮箱 ,1991年06期
- 【下载频次】17