节点文献
用氟选择性电池测定水中的氟化物
Determination of fluoride in water by means of a Fluoride - Selective Cell
【摘要】 提出了用氟化镧单晶膜作为敏感接点组成氟选择性电池测定水中氟化物的方法,建立了测定氟化物的实验条件;确定氟离子浓度的测定范围为0.2~100mg/L,相对误差小于3%,结果表明:本方法简便、实用、快速并能满足准确度要求测定水中氟化物的方法,是一种值得推荐的新方法。
【Abstract】 This paper describes application of fluoride-selective cell with a lanthanum fluoride single crystal membrane as the sensitive membrane junction to the determination method of fluoride in water. The cell consists of two half-cells in which insert two identical reference electrodes (SCE) respectively. The concentration range of dertermination of fluoride in water by direct potentiometry is 0.2-100mg/L and relative error is less 3%. This new method of determinating fluoride in water is simpler, speed and accurate. It can be widely recommended.
【关键词】 选择性电池;
氟化物;
水;
氟化镧膜;
【Key words】 Selective Cell; Fluoride; Water; lanthanam fluoride menbranc;
【Key words】 Selective Cell; Fluoride; Water; lanthanam fluoride menbranc;
- 【文献出处】 北京工业大学学报 ,Journal of Beijing Polytechnic University , 编辑部邮箱 ,1991年03期
- 【被引频次】1
- 【下载频次】52