用自制睑外眼球压迫器和非接触眼压计进行睑外眼球压迫试验的报告(正常值)
EXTRAPALPEBRAL BULBAR COMPRESSION TEST BY USING A SELF—DESIGNED EXTRAPALPEBRAL BULBAR COMPRESSOR AND NONCONTACT TONOMETER (NORMAL VALUE )
【摘要】 用自制睑外眼球压迫器压迫眼球4分钟,并用非接触眼压计测量眼压,共对162只正常眼进行了检查。结果平均下降率为32.03%,正常下降范围为32.03%±1.96×5.5%=32.03±10.78%,属正态分布。眼球压迫前后对比具有明显显著性差异。说明本法可应用于临床,作为对慢性单纯性青光眼的诊断方法之一。此法简单、安全、准确,压迫器容易制作。
【Abstract】 We have made extrapalpebral bulbar compression tests on 160 normal eyes in 81 patients using a self-designed extrapalpebral compressor. The intraocular pressure was examined by a noncontact tonometer (N C T) .The averages of the intraocular pressure before and after compression were 11.4 turn Hg and 3.46 mm Hg respectively (p<0.01) . Therefore, the difference between the intraocular pressure before and after compression was marked. The formula (P0-Pt)/(p0)×100% was used to calculate the rate of drop in the intraocular pressure. The average of the intraocular pressure drop was 32.03%. The standard deviation was 5.5%. The range of dropping was 32.3±10.78%. As this method is applied without direct contact to the cornea, it is safe and accurate. we suggest that this method may be used clinically for the detection of open-angle glaucoma.
- 【文献出处】 暨南理医学报(医学专版) , 编辑部邮箱 ,1984年02期